New Validated Stability-Indicating Rp-HPLC Method for Simultaneous Estimation of Atorvastatin and Ezetimibe in Human Plasma by Using PDA Detector

Kumar, S. Ashutosh and Debnath, Manidipa and Rao, J.V.L.N. Seshagiri and Sankar, D. Gowri (2015) New Validated Stability-Indicating Rp-HPLC Method for Simultaneous Estimation of Atorvastatin and Ezetimibe in Human Plasma by Using PDA Detector. Advanced Pharmaceutical Bulletin, 5 (3). pp. 385-391. ISSN 2228-5881

[thumbnail of APB-5-385.pdf] Text
APB-5-385.pdf - Published Version

Download (631kB)

Abstract

Purpose: This paper describes a simple, precise and accurate RP-HPLC method for simultaneous estimation of atorvastatin and ezetimibe in plasma.

Methods: The chromatographic separation of the drugs were performed on an X-Terra C8 (4.6 x 150 mm, 3.5 mm), with phosphate buffer [pH 3.5 with Ortho Phosphoric Acid] – acetonitrile 40:60 (v/v) as mobile phase. The detection was performed at 235 nm. The flow rate was maintained at 1.2 mL/min. The run time was 8.0 min.

Results: The accuracy and reliability of the method was assessed by evaluation of linearity (5-25 µg/mL for both atorvastatin calcium and ezetimibe), precision (intra-day RSD 0.57 % and inter-day RSD 0.02 % for atorvastatin calcium and intra-day RSD 0.56 % and inter-day RSD 0.1 % for ezetimibe), accuracy (100.08- 100.84 % for atorvastatin calcium and 100.56- 101.00 % for ezetimibe), and specificity, in accordance with ICH guidelines. The LLOQ obtained by the proposed method were 1.294 and 1.384 µg/mL for atorvastatin and ezetimibe respectively.

Conclusion: Overall the proposed method was found to be suitable and accurate for the quantitative determination in plasma. The method was effectively separated the drug from plasma.